No stress rentals! We help you to choose the equipment that you need, we expedite shipment, and follow up.
| MSOX6004A X-Series specification overview | |
| Bandwidth (-3dB) | 1 GHz |
| Calculated rise time (10 to 90%) | ≤ 350 ps |
| Input channels | 4 + 16 |
| Maximum sample rate | 20 GSa/s half channels, 10 GSa/s full channels |
| Maximum memory depth | 4 Mpts half channels, 2 Mpts all channels |
| Display size and type | 12.1-inch capacitive multi-touch/gesture-enabled display |
| Waveform update rate | > 450,000 waveforms per second |
| Vertical system analog channels | |
| Hardware bandwidth limits | 1 MΩ: 20 MHz, 200 MHz (selectable per channel) 50 Ω: 20 MHz, 200 MHz, 1.5 GHz, 3 GHz (selectable per channel) |
| Input coupling | AC, DC |
| Input impedance | Selectable: 1 MΩ ± 1% (14 pF), 50 Ω ± 3% |
| Input sensitivity range | 1 mV/div to 5 V/div at 1 MΩ (200 MHz bandwidth limit at ≤ 2 mV/div) 1 mV/div to 1 V/div at 50 Ω |
| Vertical resolution | 8 bits (measurement resolution is 12 bits with averaging) |
| Maximum input voltage | 1 MΩ: 30 Vrms or ± 40 Vmax (DC + Vpeak); Probing technology allows testing of higher voltages. The included N2894A 10:1 probe supports 300 Vrms or ± 400 Vmax (DC + Vpeak) 50 Ω: ± 5Vpk max |
| DC vertical gain accuracy | ± 3 div: 1.5% of full scale (typical)± 4 div: 2.5% of full scale (warranted) |
| DC vertical offset accuracy | ± 0.1 div ± 2 mV ± 1% of offset setting (valid for an offset of ≤ ± 12 divisions) |
| Channel-to-channel isolation | ≥ 100:1 (DC to 1 GHz), ≥ 30:1 (> 1 GHz) |
| Vertical system digital channels | |
| Digital input channels | 16 digital (D0 to D15. pod 1: D7 ~ D0, Pod 2: D15 ~ D8) |
| Thresholds | Threshold per pod |
| Threshold selections | TTL (+1.4 V), 5 V CMOS (+2.5 V), ECL (-1.3 V), user-defined (selectable by pod) |
| User-defined threshold range | ± 8.0 V in 10 mV steps |
| Maximum input voltage | ± 40 V peak |
| Threshold accuracy | ± (100 mV + 3% of threshold setting) |
| Maximum input dynamic range | ± 10 V about threshold |
| Minimum voltage swing | 500 mVpp |
| Input impedance | 100 kΩ ± 2% at probe tip |
| Input capacitance | ~ 8 pF |
| Vertical resolution | 1 bit |
| Display characteristics | |
| Display | 12.1-inch capacitive multi-touch/gesture enabled color TFT LCD |
| Display mode | Zone/zoom/annotation mode and waveform placement mode |
| Resolution | 800 (H) x 480 (V) pixel format (screen area) |
| Graticules | 8 vertical divisions by 10 horizontal divisions with intensity controls |
| Format | YT, XY, and Roll |
| Maximum waveform update rate | > 135,000 wfm/s (real time) > 450,000 wfm/s (real time max update rate) |
| Persistence | Off, infinite, variable persistence (100 ms to 60 s) |
| Intensity gradation | 256 intensity levels |
| General and environmental characteristics | |
| Power line consumption | Max 200 W |
| Power voltage range | 100 to 120 V, 50/60/400 Hz 100 to 240 V, 50/60 Hz |
| Environmental rating | 0 to +50 °C; 3962 m Max Maximum Relative Humidity (non-condensing): 80%RH up to 40 °C From 40 °C to 50 °C, the maximum % Relative Humidity follows the line of constant dew point |
| Safety | ANSI/UL Std. No. 61010-1:2012; CAN/CSA-C22.2 No. 61010-1-12 |
| Vibration | Meets IEC60068-2-6 and MIL-PRF-28800; class 3 random |
| Shock | Meets IEC 60068-2-27 and MIL-PRF-28800; class 3 random Operating 30 g, ½ sine. 11 ms duration, 3 shocks/axis along major axis, total of 18 shocks |
| Dimensions (W x H x D) | 425 mm W x 288 mm H x 148 mm D (16.7 x 11.3 x 5.8)in. |
| Weight | Net: 6.8 kg (15 lbs.), Shipping: 11.3 kg (25 lbs.) |
| For detailed specifications please see: Keysight MSOX6004A Data Sheet | |